Government Decree N185 establishes binding technical requirements for disinfection and sterilisation in all Georgian healthcare, public health and public-interest facilities. It defines approved methods, required concentrations and exposure times, equipment standards, validation procedures and monitoring obligations — aligned with European standards EN ISO 17665 and EN 556.
Spaulding classification — reprocessing requirements
| Spaulding category | Device type | Required reprocessing level | Examples |
|---|---|---|---|
| Critical | Enters sterile tissue, vascular system or body cavity | Sterilisation | Surgical instruments, implants, cardiac catheters, biopsy forceps |
| Semi-critical | Contacts intact mucous membranes or non-intact skin | High-level disinfection (HLD) | Endoscopes, laryngoscope blades, respiratory therapy equipment, vaginal specula |
| Non-critical | Contacts intact skin only | Low-level disinfection or cleaning | Blood pressure cuffs, stethoscopes, bed rails, patient transport chairs |
Approved sterilisation methods
| Method | Parameters | Suitable for | Validation standard |
|---|---|---|---|
| Steam sterilisation (autoclaving) | 134°C / 18 min (porous load); 121°C / 15 min (simple cycle) | Heat-stable instruments, textiles, fluids | EN ISO 17665-1 |
| Ethylene oxide (EO) | 55°C, 600 mg/L, 60% RH, minimum 3 hours | Heat/moisture-sensitive devices | EN ISO 11135 |
| Hydrogen peroxide plasma | Per manufacturer validation; typically 45–55°C | Endoscopes, electronics, heat-sensitive items | EN ISO 14937 |
| Dry heat | 170°C / 60 min or 160°C / 120 min | Glassware, metal instruments (no lumens) | EN ISO 20857 |
High-level disinfection (HLD) for endoscopes
Sterile supply unit (SSU) requirements
Healthcare facilities performing sterilisation must maintain a dedicated sterile supply unit with: physical separation of decontamination, packaging and sterile storage zones; positive pressure in clean areas; negative pressure in decontamination areas; validated autoclaves subject to daily, weekly and quarterly testing; biological indicators tested weekly; Bowie-Dick test for pre-vacuum autoclaves daily; autoclave cycle records retained for minimum 5 years; traceability of sterilised items to patient records for critical devices.
In plain language
Decree N185 sets the technical requirements for disinfection and sterilisation in healthcare and other public-interest facilities: which methods and agents to use, how equipment is validated and how processes are monitored.
What it means in practice
| Who | What changes |
|---|---|
| Patients | Instruments and devices used on patients must be safely reprocessed to prevent infection. |
| Clinicians and providers | Apply the correct level of reprocessing for each device — sterilisation, high-level or low-level disinfection. |
| Facility managers | Operate a sterile supply unit or contracted service with validated equipment, monitoring and records. |
International frame of reference
- WHO/PAHO, Decontamination and reprocessing of medical devices for health-care facilities (2016)
- WHO Guidelines on core components of IPC programmes (2016)
Questions and answers
What is the Spaulding classification?
A system that assigns the required level of reprocessing by infection risk: critical items need sterilisation, semi-critical high-level disinfection and non-critical low-level disinfection. See the table above.
Where are related protocols?
See IPC resources.
Related pages
Note. This page summarises Georgian law and regulation for information and is not legal advice. The authoritative text is the one published in the Legislative Herald of Georgia (matsne.gov.ge); where this summary and the source differ, the source prevails. Last verified: 16 October 2026. Corrections: info@accreditation.ge. Legal notice.