Disinfection and sterilisation rules: Decree N185

Georgia's Technical Regulation N185 on disinfection and sterilisation in medical, public health and public-interest facilities. Methods, concentrations, validation and monitoring requirements.

Background photo: GTour (gtour.ge)

Regulations › Legal Framework

Type
Government Technical Regulation
Reference
Government Decree N185
Year
2014
Issuing authority
Cabinet of Ministers of Georgia
Scope
All healthcare, public health and public-interest facilities in Georgia
Official source
matsne.gov.ge

Government Decree N185 establishes binding technical requirements for disinfection and sterilisation in all Georgian healthcare, public health and public-interest facilities. It defines approved methods, required concentrations and exposure times, equipment standards, validation procedures and monitoring obligations — aligned with European standards EN ISO 17665 and EN 556.

Spaulding classification — reprocessing requirements

Spaulding categoryDevice typeRequired reprocessing levelExamples
CriticalEnters sterile tissue, vascular system or body cavitySterilisationSurgical instruments, implants, cardiac catheters, biopsy forceps
Semi-criticalContacts intact mucous membranes or non-intact skinHigh-level disinfection (HLD)Endoscopes, laryngoscope blades, respiratory therapy equipment, vaginal specula
Non-criticalContacts intact skin onlyLow-level disinfection or cleaningBlood pressure cuffs, stethoscopes, bed rails, patient transport chairs

Approved sterilisation methods

MethodParametersSuitable forValidation standard
Steam sterilisation (autoclaving)134°C / 18 min (porous load); 121°C / 15 min (simple cycle)Heat-stable instruments, textiles, fluidsEN ISO 17665-1
Ethylene oxide (EO)55°C, 600 mg/L, 60% RH, minimum 3 hoursHeat/moisture-sensitive devicesEN ISO 11135
Hydrogen peroxide plasmaPer manufacturer validation; typically 45–55°CEndoscopes, electronics, heat-sensitive itemsEN ISO 14937
Dry heat170°C / 60 min or 160°C / 120 minGlassware, metal instruments (no lumens)EN ISO 20857

High-level disinfection (HLD) for endoscopes

Endoscope reprocessingAll flexible endoscopes contacting mucous membranes must undergo high-level disinfection between patients using the following sequence: (1) pre-cleaning at bedside immediately after use; (2) leak testing; (3) manual cleaning with enzymatic detergent; (4) rinsing; (5) high-level disinfection with glutaraldehyde 2% (minimum 20 minutes at 20°C) or OPA 0.55% (minimum 12 minutes at 20°C) or peracetic acid per manufacturer guidance; (6) rinsing with sterile or filtered water; (7) drying with filtered air; (8) storage in a clean, ventilated cabinet. Automated endoscope reprocessors (AERs) must be validated and subject to periodic performance qualification.

Sterile supply unit (SSU) requirements

Healthcare facilities performing sterilisation must maintain a dedicated sterile supply unit with: physical separation of decontamination, packaging and sterile storage zones; positive pressure in clean areas; negative pressure in decontamination areas; validated autoclaves subject to daily, weekly and quarterly testing; biological indicators tested weekly; Bowie-Dick test for pre-vacuum autoclaves daily; autoclave cycle records retained for minimum 5 years; traceability of sterilised items to patient records for critical devices.

In plain language

Decree N185 sets the technical requirements for disinfection and sterilisation in healthcare and other public-interest facilities: which methods and agents to use, how equipment is validated and how processes are monitored.

What it means in practice

WhoWhat changes
PatientsInstruments and devices used on patients must be safely reprocessed to prevent infection.
Clinicians and providersApply the correct level of reprocessing for each device — sterilisation, high-level or low-level disinfection.
Facility managersOperate a sterile supply unit or contracted service with validated equipment, monitoring and records.

International frame of reference

Questions and answers

What is the Spaulding classification?

A system that assigns the required level of reprocessing by infection risk: critical items need sterilisation, semi-critical high-level disinfection and non-critical low-level disinfection. See the table above.

Where are related protocols?

See IPC resources.

Note. This page summarises Georgian law and regulation for information and is not legal advice. The authoritative text is the one published in the Legislative Herald of Georgia (matsne.gov.ge); where this summary and the source differ, the source prevails. Last verified: 16 October 2026. Corrections: info@accreditation.ge. Legal notice.